Long Story Short: 125 Years of Quality.
Celebrate with us and get your anniversary discount.
Compare
PG 8504 [ADP]
Laboratory dishwasher with feed pump for AD water.
- Perf./cycle, e.g. 64 narrow necked/lab glassware w/ injectorsFurther stacking options
At the same time it is possible to reprocess other laboratory glassware in the upper basket. - Efficient use of resources - variable speed heater pumpThe heart of perfect results
In the individual programmes the speed of the pump and thus the wash pressure is adjusted in line with the relevant requirements – high pressure facilitates the removal of stubborn soiling, whilst low pressure guarantees the complete wetting of all the wash items. As a result, the consumption of water and energy remains particularly low with optimum wash pressure at all times. - Dispense process chemicals easily: Powder dispenser in the door
- Pump for DI water: final rinse with fully demineralised water
Product benefits
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Clear text display
Text display
Everything at a glance: The display shows everything as clear text in the country-specific language.
Large selection of accessories
Reprocess flexibly
Flexible and comprehensive components for optimum cleaning performance and a variety of reprocessing options.
Can be built under a worktop
For a consistent look
The machine can be easily integrated under a worktop.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Product details
PG 8504 [ADP]
EAN 4002515565285 / Article number 62850422GB / Material number 10096790
EAN 4002515565285 / Article number 62850422GB / Material number 10096790
Construction type | Freestanding, can be built-under, slimline |
---|---|
Casing | Stainless steel |
Can be built-under | |
Electric door lock | |
Buzzer, acoustic signal at end of programme | |
Service-friendly design |
Narrow-necked glassware per cycle [number] | 64 |
---|
Circulation pump, Qmax in l/min | 500 |
---|---|
Maximum final rinse temperature in °C | 70 |
Wash cabinet usable capacity in l | 145 |
Tested operating hours Note Our machines undergo intensive testing under laboratory conditions. The actual operating hours may vary depending on actual usage behaviour, maintenance and care of the machine. | 15000 |
Control type | TouchControl |
---|---|
Programme selection | Touch on MSetal Touch on Metal Greater user convenience and easy surface cleaning. |
Direct-access pushbuttons | |
Max. delay start in h | 24 |
Programmability | programmable |
Programmes [number] | 6 |
Free programme positions [number] | 2 |
Automatic programme restart | Programme failure protection Time-saving and safe: The programme failure protection facilitates the continuation of the cleaning and disinfection process after a short interruption, such as a power cut. If the success of the reprocessing process should be at risk, the appliance issues a corresponding warning message. |
Time left display | |
Programme sequence display | |
Selectable display languages |
Electrical connection | AC 230V 50HZ |
---|---|
Electric heater rating in kW | 1.7 |
Total rated load in kW | 2.4 |
Fuse rating in A | 13 |
Length of mains connection cable in m Note The length refers to the supply lead on the outside of the appliance. | 1.9 |
Cold water [number] | 1 |
---|---|
VEP – with feed pump for DI water [number] | 1 |
Required flow pressure in kPa Note on ADP variants This value applies to cold, hot and DI water; however, it does not apply to DI water intake on ADP variants because these use a feed pump. | 200-1000 |
Maximum delivery head for drain pump in cm | 100 |
Integrated water softener | |
Maximum water hardness (cold water/hot water) in mmol/l | 10.7 |
Drain pump [DN] | 22 |
Water protection system Waterproof System (WPS) Protection against flooding: In the event of problems, the system stops water intake automatically. | Waterproof system |
External dimensions, net height in mm | 835 |
---|---|
External dimensions, net width in mm | 600 |
External dimensions, net depth in mm | 600 |
External dimensions, gross height in mm Note Including packaging | 940 |
External dimensions, gross width in mm Note Including packaging | 675 |
External dimensions, gross depth in mm Note Including packaging | 745 |
Wash cabinet, usable height in mm | 520 |
Wash cabinet, usable width in mm | 530 |
Wash cabinet, depth of upper basket in mm | 474 |
Wash cabinet, depth of lower basket in mm | 520 |
Net weight in kg | 72 |
Gross weight in kg Note Including packaging | 79 |
Maximum floor load in N/m² Maximum floor bearing capacity Maximum floor bearing capacity states the maximum weight (N) that can be placed on the supporting surface on which the machine is installed. This maximum load bearing capacity is the result of adding up the weight of the machine, the load and the water in it, as well as the dynamic forces that occur during operation. | 1200 |
Emission sound pressure level in the workplace Emissions - sound pressure at workplace The emission sound pressure level depends on the programme and is a value given to the noise created by the machine. Miele Professional appliances have very low noise levels. | <lt/>70 dB(A) re 20 µPa |
---|---|
Heat dissipation into the room in MJ/h Heat output into the room During operation warm air is vented into the surrounding room. The released amount of heat/time depends on the set temperatures and the intensity of use. The stated maximum value can be used to control the temperature of the area or to calculate an increase in room temperature. | 1.1 |
Short | |
---|---|
Medium | |
Long |
External dispensing modules |
---|
VDE (electrical safety) | |
---|---|
IP safety class according to EN 60529: IP 21 | |
EN 61010-1 | |
EN 61010-2-040 | |
RoHS guideline 2011/65/EU | |
WRAS | |
Machinery Directive 2006/42/EC |
Additional accessories and consumables
Find additional accessories and consumables to go with this product
Additional accessories for laboratory technology
Find here the accessories that can be bought additionally for this product.
Additional consumables for laboratory technology
Find here the consumables that can be bought additionally for this product.